Beschreibung der Beschaffung
The SQUID magnetometer must also meet the following requirements:
- Superconduction DC magnet, wet system
- Magnetic field shielding (superconductive and/or soft magnetic)
- Magnetic field strength higher or equal than 5 Tesla
- Measurement of both longitudinal (along cylinder z axis) and horizontal magnetic moment (along x and y axis).
- Liquid helium hold time at any operation condition: t>72 hrs
- Liquid nitrogen hold time at any operation condition: t>72 hrs
- Temperature stability: at least 0.1K over t ≥ 10 min
- Magnetic field degauss function implemented
- Real-time remote control, measurement sequence operation and data display software
- Sample temperature range: 2 - 300 K
- Sample space diameter: at least 8 mm
- Sample top loading system for easy sample scan / movement / rotation
- AC measurement modus integrated
- delivery has to took place 12 month (=52 weeks) after placing the order
- site acceptance test has to take place within 1 month (=4 weeks) after delivery
- The delivery has to be completed not later than 12 months (=52 weeks) after the oder is placed
- delivery at the place of usage according to DAP/DDP Dresden (Incoterms 2020)
(Place of Usage: HLD, Building 550/551, Bautzner Landstraße 400, 01328 Dresden)
- installation and instruction at place
- Site Acceptance Test
- user documentation in English or German
- 1y warranty, software upgrades, remote support